Plain-language terms, and the definition of every number the simulator shows.

Terms

Preload · also end-diastolic volume as preload
How far the ventricle is stretched at the end of filling, best described by end-diastolic volume. Filling pressure is related to it but is not the same thing.
Afterload
The load the ventricle ejects against: arterial pressure, vascular resistance and compliance together.
Contractility · also inotropy
How strongly the muscle can generate pressure at a given volume. In the model it is the end-systolic elastance, Ees.
Ees · also end-systolic elastance
Slope of the end-systolic pressure–volume line. Higher means stiffer, stronger contraction. Exact definition
Ea · also arterial elastance
Effective arterial elastance: end-systolic pressure divided by stroke volume, a lumped description of afterload. Exact definition
ESPVR · also end-systolic pressure–volume relationship
The line joining end-systolic points of loops made under different loading.
EDPVR · also end-diastolic pressure–volume relationship
The curve of passive filling pressure against volume. It steepens as the ventricle fills, so the last mL of filling costs the most pressure.
LVEDV · also EDV, end-diastolic volume
Left ventricular volume at the end of filling. The best simple picture of preload. Exact definition
LVESV · also ESV, end-systolic volume
Left ventricular volume at the end of ejection. Exact definition
Stroke volume · also SV, LV SV
Blood ejected per beat: end-diastolic minus end-systolic volume (net of any regurgitation). Exact definition
Ejection fraction · also EF, LVEF
Stroke volume as a percentage of end-diastolic volume. It depends on loading as well as contractility. Exact definition
Filling pressure (LVEDP) · also LVEDP, filling pressure
Left ventricular pressure at the start of ventricular contraction: the pressure cost of the volume the ventricle holds. Exact definition
Wedge pressure (PAWP) · also PAWP, PCWP, wedge
Pressure recorded with a balloon catheter wedged in a pulmonary artery branch, an estimate of left atrial pressure when the wedge is complete. Exact definition
Right atrial pressure (RAP) · also RAP, CVP, central venous pressure
Mean pressure in the right atrium, a downstream pressure for venous return. Exact definition
Mean arterial pressure (MAP) · also MAP
Average aortic pressure over the beat. Exact definition
Cardiac output · also CO
Net flow per minute into the systemic arteries: stroke volume times heart rate. Exact definition
Cardiac index · also CI
Cardiac output divided by body surface area. Exact definition
SVR · also systemic vascular resistance
Pressure drop across the systemic circulation divided by flow. A calculated number, not a measurement of the arterioles themselves. Exact definition
PVR · also pulmonary vascular resistance
Pressure drop across the lungs divided by flow. Exact definition
Stressed volume · also unstressed volume
The part of blood volume that stretches the vessel walls and generates pressure. Unstressed volume only fills the space.
Mean systemic filling pressure · also MSFP
The pressure in the systemic circulation if the heart stopped and flow equalised. It drives venous return.
Venous return
Flow from the systemic veins back to the right atrium. At steady state it equals cardiac output.
Frank–Starling relationship · also Starling
A fuller ventricle ejects a larger stroke volume, up to a point.
Elastance · also time-varying elastance
Pressure per unit of volume above the chamber's unstressed volume. It rises and falls through each beat.
Compliance
Volume change per unit of pressure change: the inverse of stiffness.
Transmural pressure · also distending pressure
Pressure inside a chamber minus the pressure outside it. This is what stretches the chamber, not the measured pressure alone.
Pulse pressure variation · also PPV, SVV
Breath-to-breath swing in pulse pressure on a ventilator. It is only interpretable under specific conditions.
Qp/Qs · also shunt ratio
Pulmonary flow divided by systemic flow. Above 1 means net left-to-right shunting. Exact definition
Fick method · also thermodilution
Cardiac output from oxygen consumption divided by the arteriovenous oxygen difference.
Baseline
The state you compare against. Here it is the starting patient and stays fixed until you replace it.

Measurements

Rhythm

Heart rate (bpm)
60 / RR interval of the displayed beat.
  • Instantaneous rate of one beat; irregular rhythms vary beat to beat.
Model-derived (not directly measurable in a patient) · validity: Always defined.

Left ventricle

LV end-diastolic volume (mL)
Maximum LV volume in the beat.
  • Chamber volume is a model state, not an imaging estimate.
Model-derived (not directly measurable in a patient) · validity: Always defined.
LV end-systolic volume (mL)
Minimum LV volume in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
LV total stroke volume (mL)
LVEDV − LVESV. Total chamber excursion, which includes any regurgitant volume.
  • Not equal to forward stroke volume when the aortic or mitral valve leaks.
Calculated from signals · validity: Always defined; compare with lv.forwardSv in regurgitation.
LV ejection fraction (%)
100 × (LVEDV − LVESV) / LVEDV.
  • Total ejection fraction, not effective forward EF.
  • EF is load dependent and is not a measure of contractility (Ees).
Calculated from signals · validity: Undefined if LVEDV ≤ 0.

Flow and output

Forward aortic stroke volume (mL)
Net aortic valve flow per beat: ejected volume minus aortic regurgitant volume (effective forward stroke volume).
  • Net flow through the aortic valve only; device flow is reported separately.
Calculated from signals · validity: Always defined.
Volume ejected through the aortic valve (mL)
Integral of positive aortic valve flow. Exceeds the effective forward volume in aortic regurgitation.Calculated from signals · validity: Always defined.
Forward pulmonic stroke volume (mL)
Net pulmonic valve flow per beat (ejected minus pulmonic regurgitant volume).Calculated from signals · validity: Always defined.
Systemic cardiac output (L/min)
Effective forward (net aortic) stroke volume × heart rate / 1000.
  • In an unsupported circulation without shunts this equals systemic flow.
  • During MCS, device flow is reported separately and not added here.
Calculated from signals · validity: Always defined.
Pulmonary blood flow (L/min)
Net pulmonic stroke volume × heart rate / 1000.
  • Equals systemic flow in steady state without shunts.
Calculated from signals · validity: Always defined.
Mitral regurgitant volume (mL)
Integral of negative (retrograde) mitral flow over the beat.Calculated from signals · validity: Reported as 0 when the valve is competent.
Aortic regurgitant volume (mL)
Integral of negative (retrograde) aortic valve flow over the beat.Calculated from signals · validity: Reported as 0 when the valve is competent.
Tricuspid regurgitant volume (mL)
Integral of negative tricuspid flow over the beat.Calculated from signals · validity: Reported as 0 when the valve is competent.
Mitral regurgitant fraction (%)
100 × MR volume / total LV stroke volume.
  • Model flow fraction; not a clinical severity grade without a reviewed mapping.
Calculated from signals · validity: Not applicable when the valve is competent.
Aortic regurgitant fraction (%)
100 × AR volume / volume ejected through the aortic valve.
  • Model flow fraction; not a clinical severity grade without a reviewed mapping.
Calculated from signals · validity: Not applicable when the valve is competent.
Native aortic output (L/min)
Net aortic valve flow × heart rate / 1000 (blood ejected by the LV through its own valve).
  • Excludes device flow; see co.impella, co.ecmo and co.total.
Calculated from signals · validity: Always defined.
Microaxial pump flow (L/min)
Mean LV-to-aorta pump flow over the beat (parallel to the aortic valve).
  • Pressure–flow characteristic is a placeholder until manufacturer curves are mapped (device record).
Model-derived (not directly measurable in a patient) · validity: Not applicable when the pump is off.
VA-ECMO circuit flow (L/min)
Mean RA-to-arterial circuit flow over the beat (parallel to the whole heart).
  • Set flow reduced by arterial afterload sensitivity and RA drainage limits (placeholder circuit).
Model-derived (not directly measurable in a patient) · validity: Not applicable when the circuit is off.
Total systemic output (L/min)
Native aortic + microaxial + ECMO flow: these are parallel paths into the systemic arteries, so they add. Serial segments are never added.
  • Cross-checked against mean systemic arteriolar flow (co.systemicCheck).
Calculated from signals · validity: Always defined.
Mean systemic arteriolar flow (conservation check) (L/min)
Mean flow through the systemic arterioles; equals total output at steady state if the circuit conserves flow.Model-derived (not directly measurable in a patient) · validity: Always defined.
Aortic valve open fraction (%)
Percentage of the beat during which the aortic valve is open (0 = the valve never opens, as under high VA-ECMO flow).Model-derived (not directly measurable in a patient) · validity: Always defined.
Pump suction present (% of beat)
Percentage of the beat in which microaxial inflow is limited by a small LV cavity.
  • Suction is modelled as a smooth flow limit below the suction volume; no alarm logic.
Model-derived (not directly measurable in a patient) · validity: Not applicable when the pump is off.
Pulmonary-to-systemic flow ratio (unitless)
Net pulmonic flow divided by net systemic (aortic) flow.
  • Equals 1 without shunts; > 1 for left-to-right shunting.
Calculated from signals · validity: Undefined if systemic flow ≈ 0.
Net shunt flow (ASD + VSD, left-to-right positive) (L/min)
Mean net flow across the modelled defects per minute.Model-derived (not directly measurable in a patient) · validity: Zero without defects.

Pressures

LV end-diastolic pressure (mmHg)
LV pressure at the onset of ventricular activation (QRS-equivalent).
  • End-diastole defined by activation onset; in aortic regurgitation this precedes the maximal-volume sample.
Model-derived (not directly measurable in a patient) · validity: Always defined.
LV peak systolic pressure (mmHg)
Maximum LV pressure in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
Aortic systolic pressure (mmHg)
Maximum central aortic root pressure in the beat.
  • Central aortic, not peripheral cuff or radial pressure.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Aortic diastolic pressure (mmHg)
Minimum central aortic root pressure in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
Mean arterial pressure (mmHg)
Time average of central aortic root pressure over the beat.
  • True time average, not the 1/3 pulse pressure estimate.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Aortic pulse pressure (mmHg)
Ao sys − Ao dia.Calculated from signals · validity: Always defined.
PA systolic pressure (mmHg)
Maximum proximal pulmonary artery pressure in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
PA diastolic pressure (mmHg)
Minimum proximal pulmonary artery pressure in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
Mean PA pressure (mmHg)
Time average of proximal PA pressure over the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
Mean pulmonary artery wedge pressure (model) (mmHg)
Time average of the pulmonary venous compartment pressure. Stands in for the wedge pressure a catheter would record; the observation layer (M11) adds catheter artifacts separately.
  • Equals mean pulmonary venous pressure in the model; a real wedge includes catheter position and damping effects.
  • No respiratory gating applied (respiratory swing is zero unless enabled).
Model-derived (not directly measurable in a patient) · validity: Always defined in the model; measurement reliability is a separate question (M11).
Mean left atrial pressure (mmHg)
Time average of LA pressure over the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
Mean right atrial pressure (mmHg)
Time average of RA pressure over the beat. Equivalent to CVP in the model.
  • No respiratory gating applied.
Model-derived (not directly measurable in a patient) · validity: Always defined.
LV–aortic instantaneous peak gradient (mmHg)
Maximum of (LV − aortic root pressure) while the aortic valve is open.
  • True-signal gradient; the cath lab recomputes it from recorded signals.
  • Includes any dynamic LVOT gradient.
Model-derived (not directly measurable in a patient) · validity: Not applicable if the valve never opens.
LV–aortic mean gradient (mmHg)
Mean of (LV − aortic root pressure) over the ejection period.
  • Gradient rises with flow² for a fixed orifice.
Model-derived (not directly measurable in a patient) · validity: Not applicable if the valve never opens.
Transmitral mean diastolic gradient (mmHg)
Mean of (LA − LV pressure) while the mitral valve is open.
  • Rises with flow and with shorter diastole (rate).
Model-derived (not directly measurable in a patient) · validity: Not applicable if the valve never opens.
Transtricuspid mean diastolic gradient (mmHg)
Mean of (RA − RV pressure) while the tricuspid valve is open.Model-derived (not directly measurable in a patient) · validity: Not applicable if the valve never opens.
Dynamic LVOT peak gradient (mmHg)
Maximum pressure drop across the dynamic outflow-tract orifice (area falls as the LV empties).
  • Volume-dependent orifice model; provocation changes it through preload, afterload and contractility.
Model-derived (not directly measurable in a patient) · validity: Not applicable unless dynamic obstruction is enabled.
LA v wave (peak) (mmHg)
Peak LA pressure while the mitral valve is closed.
  • A large v wave is not a stand-alone MR severity classifier; it depends on LA compliance.
Model-derived (not directly measurable in a patient) · validity: Always defined.
LA a wave (peak) (mmHg)
Peak LA pressure in the 150 ms before mitral closure (atrial contraction).Model-derived (not directly measurable in a patient) · validity: Not applicable without organised atrial contraction.
RA v wave (peak) (mmHg)
Peak RA pressure while the tricuspid valve is closed.
  • Prominent in TR; also raised by cannon waves in AV dissociation.
Model-derived (not directly measurable in a patient) · validity: Always defined.
RA a wave (peak) (mmHg)
Peak RA pressure in the 150 ms before tricuspid closure.Model-derived (not directly measurable in a patient) · validity: Not applicable without organised atrial contraction.
Mean pericardial pressure (model) (mmHg)
Time average of the modelled pericardial pressure.
  • Exponential pericardial relation of total heart volume plus effusion.
Model-derived (not directly measurable in a patient) · validity: Not applicable when the pericardium is disabled.
LV end-diastolic transmural pressure (mmHg)
LV pressure minus external (transmitted pleural + pericardial) pressure at end-diastole: the pressure that actually distends the chamber.
  • Differs from the measured LVEDP whenever pleural or pericardial pressure is not zero.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Mean RA transmural pressure (mmHg)
Mean RA pressure minus transmitted pleural and pericardial pressure.Model-derived (not directly measurable in a patient) · validity: Always defined.
IABP diastolic augmentation (mmHg)
Peak aortic pressure while the balloon is inflated minus the unassisted systolic peak.
  • Depends on balloon volume, arterial compliance and timing.
Model-derived (not directly measurable in a patient) · validity: Not applicable when the balloon is off or not inflated in this beat.
Assisted aortic end-diastolic pressure (mmHg)
Aortic pressure at ventricular activation onset (the pressure the LV must exceed to eject).Model-derived (not directly measurable in a patient) · validity: Always defined.

Right ventricle

RV end-diastolic volume (mL)
Maximum RV volume in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
RV end-systolic volume (mL)
Minimum RV volume in the beat.Model-derived (not directly measurable in a patient) · validity: Always defined.
RV total stroke volume (mL)
RVEDV − RVESV.
  • Includes regurgitant volume when the tricuspid or pulmonic valve leaks.
Calculated from signals · validity: Always defined.
RV ejection fraction (%)
100 × (RVEDV − RVESV) / RVEDV.
  • Load dependent; not contractility.
Calculated from signals · validity: Undefined if RVEDV ≤ 0.

Indices

Cardiac index (L/min/m²)
CO / body surface area.
  • BSA is a scenario parameter.
Calculated from signals · validity: Undefined if BSA ≤ 0.
Cardiac power output (W)
MAP × CO / 451.
  • No universal threshold is implied.
Calculated from signals · validity: Always defined.
Pulmonary artery pulsatility index (unitless)
(PA sys − PA dia) / RAP.
  • Context dependent; no universal threshold implied.
Calculated from signals · validity: Not calculated when RAP < 1 mmHg (unstable ratio).

Resistance

Systemic vascular resistance (WU)
(MAP − RAP) / CO in Wood units (mmHg·min/L). Multiply by 80 for dyn·s·cm⁻⁵.
  • Uses systemic forward flow.
Calculated from signals · validity: Not calculated when CO < 0.5 L/min.
SVR (dyn·s·cm⁻⁵) (dyn·s·cm⁻⁵)
SVR in Wood units × 80.Calculated from signals · validity: Same as SVR.
Pulmonary vascular resistance (WU)
(mPAP − PAWP) / pulmonary flow in Wood units.
  • Uses pulmonary (not systemic) flow and the model wedge pressure.
Calculated from signals · validity: Not calculated when pulmonary flow < 0.5 L/min.

Coupling

LV end-systolic pressure (mmHg)
LV pressure at aortic valve closure (last open→closed transition); falls back to the minimum-volume sample if the valve never opened.
  • End-systole defined by aortic valve closure.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Effective arterial elastance (mmHg/mL)
LVESP / total LV stroke volume.
  • Ea is a lumped afterload index, not SVR.
  • Standard unsupported-circulation definition; not valid with major regurgitation or MCS.
Calculated from signals · validity: Qualified when any valve regurgitates or a device is active; undefined if SV ≤ 0.
LV end-systolic elastance (model parameter) (mmHg/mL)
The model's contractility parameter. Known exactly because it is set, not identified from a single loop.
  • A single ordinary loop cannot identify Ees; this value is the model parameter.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Ventricular-arterial coupling ratio (unitless)
Ea / Ees using the model Ees.
  • Interpretation assumes an unsupported circulation without major regurgitation.
Calculated from signals · validity: Qualified when Ea is qualified.

Work

LV stroke work (J)
Net loop integral ∮P dV of the LV pressure–volume trajectory; 1 mmHg·mL = 1.333×10⁻⁴ J.
  • Counter-clockwise loop positive. For a transient beat the loop is closed artificially and the value is an estimate.
Calculated from signals · validity: Always defined; labelled estimate for transient beats.
RV stroke work (J)
Net loop integral ∮P dV of the RV pressure–volume trajectory.
  • Same conventions as LV stroke work.
Calculated from signals · validity: Always defined.
LV pressure–volume area (model estimate) (J)
Stroke work plus the potential energy between the model ESPVR and EDPVR from V0 to ESV.
  • Uses the model-known relations; not identifiable from one clinical loop.
  • No myocardial oxygen consumption or efficiency is derived from it.
Model-derived (not directly measurable in a patient) · validity: Always qualified (teaching estimate).

Oxygen

Pulmonary artery saturation (mixing teaching model) (%)
Flow-weighted mixture of systemic venous blood (Fick SvO₂ on systemic flow) and left-to-right shunted arterial blood.
  • Single systemic venous saturation (no SVC/IVC difference); right-to-left shunting lowers arterial saturation, which is not fed back to SaO₂.
Calculated from signals · validity: Not applicable when the Fick balance has no solution.
Oxygen saturation step-up (PA minus systemic venous) (% points)
PA saturation minus systemic venous saturation; positive with left-to-right shunting.
  • No threshold for significance is applied without its source.
Calculated from signals · validity: Not applicable without the Fick balance.
Arterial oxygen content (mL/dL)
1.34 × Hb × SaO₂ (dissolved oxygen neglected).
  • Dissolved O₂ (0.003 × PaO₂) omitted; Hb and SaO₂ are scenario inputs.
Calculated from signals · validity: Always defined.
Oxygen delivery (mL/min)
CaO₂ × CO × 10.
  • Uses systemic forward flow of the displayed beat.
Calculated from signals · validity: Always defined.
Mixed venous oxygen saturation (steady-state Fick) (%)
SaO₂ − VO₂ / (1.34 × Hb × CO × 10), i.e. the saturation at which the Fick principle balances consumption against delivery.
  • Steady-state oxygen balance with a single systemic compartment; no regional extraction differences, so ScvO₂ is NOT derived from it.
  • VO₂ is a scenario input, not measured; organ injury, lactate and urine output are not deterministic readouts of this value.
Calculated from signals · validity: Not applicable when delivery cannot meet consumption (the balance has no solution ≥ 0); flagged as supply-limited.
Oxygen extraction ratio (%)
100 × VO₂ / DO₂.
  • Same steady-state assumptions as SvO₂.
Calculated from signals · validity: Flagged when > 100% (supply-limited).
Central venous oxygen saturation (%)
Would require a regional mixing model (SVC vs IVC extraction); not derived from SvO₂.
  • Kept distinct from SvO₂ by design (spec M12).
Measured · validity: Not available in this model version.

Volume

Total blood volume (model) (mL)
Sum of all compartment volumes at the beat start.
  • Conserved except for explicit fluid or blood-loss interventions.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Volume exchanged with external reservoir (mL)
Net volume that left the circulation into clamp reservoirs during the beat (positive = removed). Zero unless a clamp is active.
  • Only non-zero in isolated-physiology mode; shows that volume is not conserved while clamps are active.
Model-derived (not directly measurable in a patient) · validity: Always defined.
Stressed systemic venous volume (mL)
Mean systemic venous volume − unstressed volume.
  • Model construct; not clinically measurable directly.
Model-derived (not directly measurable in a patient) · validity: Always defined.

See a number in the simulator? Open the full laboratory and choose All measurements to see it for the displayed beat.